Application of the new Monte Carlo code AlfaMC to the calibration of alphaparticle sources

Repositorio Dspace/Manakin

español português english

Application of the new Monte Carlo code AlfaMC to the calibration of alphaparticle sources

Mostrar el registro completo del ítem

Título: Application of the new Monte Carlo code AlfaMC to the calibration of alphaparticle sources
Autor: Jurado Vargas, Miguel; Fernández Timón, Alfonso; García Orellana, Carlos
Resumen: Las mediciones de partículas α fuentes requieren correcciones en la tasa de escrutinio debido a la dispersión y la auto-absorción en la fuente y el apoyo material. En este estudio, se describe un procedimiento sencillo para estimar estas correcciones con el nuevo código de Monte Carlo AlfaMC para considerar los efectos de dispersión y de auto-absorción conjuntamente, y así determinar la actividad de la α de emisores. El procedimiento propuesto se aplicó a 235 UO2 fuentes depositados en platino pulido respaldos. En general, la dependencia de la eficiencia con espesor de origen estaba en buen acuerdo con un simple considerando un modelo lineal y un comportamiento hiperbólico para fuentes delgadas y gruesas, respectivamente, aunque desviaciones significativas de este modelo fueron encontrados por fuentes muy delgada. Para estas fuentes muy delgada, la simulación Monte Carlo se reveló como un método requerido en la calibración primaria de partículas α fuentes. La eficacia de los resultados obtenidos por simulación con AlfaMC estuvieron de acuerdo con los datos disponibles de eficacia.Measurements of α-particle sources require corrections to the counting rate due to scattering and self-absorption in the source and the backing material. In this study, we describe a simple procedure to estimate these corrections using the new Monte Carlo code AlfaMC to consider the effects of scattering and self-absorption conjointly, and so to determine the activity of α emitters. The procedure proposed was applied to 235UO2 sources deposited on highly polished platinum backings. In general, the dependence of the efficiency with source thickness was in good agreement with a simple model considering a linear and a hyperbolic behavior for thin and thick sources, respectively, although significant deviations from this model were found for very thin sources. For these very thin sources, the Monte Carlo simulation revealed to be as a required method in the primary calibration of α -particle sources. The efficiency results obtained by simulation with AlfaMC were in agreement with available efficiency data.
URI: http://hdl.handle.net/10662/7624
Fecha: 2015


Ficheros en el ítem

Ficheros Tamaño Formato Ver
nuka_2015_0113.pdf 182.7Kb PDF Thumbnail

El ítem tiene asociados los siguientes ficheros de licencia:

Este ítem aparece en la(s) siguiente(s) colección(ones)

Mostrar el registro completo del ítem

Atribución-NoComercial-SinDerivadas 3.0 España Excepto si se señala otra cosa, la licencia del ítem se describe como Atribución-NoComercial-SinDerivadas 3.0 España

Buscar en Mi Dehesa


Listar

Mi cuenta

Estadísticas

Ayuda

Redes sociales