Identificador persistente para citar o vincular este elemento: http://hdl.handle.net/10662/14453
Títulos: QardioArm upper arm blood pressure monitor against Omron M3 upper arm blood pressure monitor in patients with chronic kidney disease: A validation study according to the European Society of Hypertension international protocol revision 2010
Autores/as: Mazoteras Pardo, Victoria
Becerro de Bengoa Vallejo, Ricardo
Losa Iglesias, Marta Elena
López López, Daniel
Rodríguez Sanz, David
Casado Hernández, Israel
Calvo Lobo, César
Palomo López, Patricia
Palabras clave: Blood pressure;Presión arterial;Hypertension;Hipertensión;Kidney disease;Enfermedad del riñon;Mobile apps;Aplicaciones móviles;Software validation;Validación de software
Fecha de publicación: 2019
Editor/a: JMIR Publications
Resumen: Background: Hypertension is considered as a main risk factor for chronic kidney disease development and progression. Thus, the control and evaluation of this disease with new software and devices are especially important in patients who suffer from chronic kidney disease. Objective: This study aimed to validate the QardioArm mobile device, which is used for blood pressure (BP) self-measurement in patients who suffer from chronic kidney disease, by following the European Society of Hypertension International Protocol 2 (ESH-IP2) guidelines. Methods: A validation study was carried out by following the ESH-IP2 guidelines. A sample of 33 patients with chronic kidney disease self-measured their BP by using the QardioArm and Omron M3 Intellisense devices. Heart rate (HR), diastolic BP, and systolic BP were measured. Results: The QardioArm fulfilled the ESH-IP2 validation criteria in patients who suffered from chronic kidney disease. Conclusions: Thus, this study is considered as the first validation using a wireless upper arm oscillometric device connected to an app to measure BP and HR meeting the ESH-IP2 requirements in patients who suffer from chronic kidney disease. New validation studies following the ESH-IP2 guidelines should be carried out using different BP devices in patients with specific diseases.
URI: http://hdl.handle.net/10662/14453
Colección:DENFE - Artículos

Archivos
Archivo Descripción TamañoFormato 
14686.pdf187,39 kBAdobe PDFDescargar


Este elemento está sujeto a una licencia Licencia Creative Commons Creative Commons